Giorgio Bormac

ICF 400 NEW- forced air Incubator

Code: 41200832

Categories: Ovens and Incubators

Availability:

ICF 400 NEW- forced air Incubator

Features

Working range: +5°C Ta ... +80°C

LCD display

Data download and recording via USB key

8 storable programmes with 8 steps each

Single-step basic programme

Multi-step programme with up to 64 steps

130°C sterilisation programme

Other features

ArgoLab forced ventilation incubator.

Simultaneous display of all operating parameters.

Operator-silenced visual and acoustic alarms.

Digital timer 1 min - 99 h 59 min and continuous operation.

Delayed start' function of the heating cycle.

Safe temperature' function: working temperature limitation for sample protection.

Special programme for sterilisation at 130°C.

Fan adjustable in 3 levels: High = 100% - Medium = 75% - Low = 50%.

Minimum heating times and low heating power for low energy consumption in every application.

Working data logging of up to 200.000 values and download to USB stick.

Safety class 3.1 (DIN 12880): maximum temperature controller with fluid expansion.

Observation of samples without heat loss: double door (glass interior).

Through-hole Ø 5 mm as standard (upper side) for insertion of an external sensor for internal temperature control and/or certification.

Stainless steel chamber with rounded corners, fully removable anti-tip supports and shelves for easy cleaning.

Door switch for easy sample loading and unloading.

Power supply 220 ± 10% V - 50/ 60 Hz.

Supplied as standard

steel grid shelves (n°2 for models up to 55 liters, n°3 for larger models), calibration report at 37°C performed with Accredia-referenced sample instrument, USB stick for data recording and downloading.

Description

Features

Working range: +5°C Ta ... +80°C

LCD display

Data download and recording via USB key

8 storable programmes with 8 steps each

Single-step basic programme

Multi-step programme with up to 64 steps

130°C sterilisation programme

Other features

ArgoLab forced ventilation incubator.

Simultaneous display of all operating parameters.

Operator-silenced visual and acoustic alarms.

Digital timer 1 min - 99 h 59 min and continuous operation.

Delayed start' function of the heating cycle.

Safe temperature' function: working temperature limitation for sample protection.

Special programme for sterilisation at 130°C.

Fan adjustable in 3 levels: High = 100% - Medium = 75% - Low = 50%.

Minimum heating times and low heating power for low energy consumption in every application.

Working data logging of up to 200.000 values and download to USB stick.

Safety class 3.1 (DIN 12880): maximum temperature controller with fluid expansion.

Observation of samples without heat loss: double door (glass interior).

Through-hole Ø 5 mm as standard (upper side) for insertion of an external sensor for internal temperature control and/or certification.

Stainless steel chamber with rounded corners, fully removable anti-tip supports and shelves for easy cleaning.

Door switch for easy sample loading and unloading.

Power supply 220 ± 10% V - 50/ 60 Hz.

Supplied as standard

steel grid shelves (n°2 for models up to 55 liters, n°3 for larger models), calibration report at 37°C performed with Accredia-referenced sample instrument, USB stick for data recording and downloading.

Technical details
Features
Capacity
400
Maximum shelf load
20
Security class
3.1
Through-hole diameter
5
Dimension
1285 x 820 x 1060
Minimum distance between shelves
50
Frequency
50 - 60
Warranty
2
Working method
Forced ventilation
Weight
160
Absorbed power
1500
Programs
8 programs with up to 8 steps each, 1 single-step program, and 1 multi-step program with up to 64 steps
Temperature resolution
0.1
Data transfer
Yes, via USB stick
Heating time (37°C)
55
Timer
99 : 59
Ventilation
3 speeds (50 %, 75 %, 100 %)
Temperature variation at 37 °C
±0,3
Useful volume
400
Power supply
230
Internal dimensions
1000 x 500 x 800
Overtemperature security
Electronic and electromechanical
Working temperature
+5 Ta…80
Display
LCD
Regulations
2006/95/EC -2004/108/EC - EN61326-1:2006 - EN61010-1:2010 - EN61010-2-010:2003
Number of shelves
3
Temperature homogeneity at 37 °C
±0,5
Packaging dimensions (W x D x H)
1350 x 900 x 1300
Chamber sterilisation cycle
Yes, at 130°C (max 10 minutes)
Temperature ramps
No
Multi user
No
Alarm History
No
Operations history
No
Internal data logger
Yes
Maximum number of data
2167
Graph
No
Door switch
Yes
Multi-step program
Yes
Maximum number of shelves
10
Download area
Information

Request more information about this product.